文字起こし 🔗
WEBVTT
00:00:01.600 --> 00:00:13.839
Welcome to Compliance Espresso, the podcast where we break down European medical device legislation with clarity, rigor, and just a hint of irony.
00:00:14.320 --> 00:00:22.640
Short episodes designed to give you reliable guidance in just a few minutes, like a good cup of coffee.
00:00:22.879 --> 00:00:24.480
I'm Nicola Focci.
00:00:24.640 --> 00:00:32.719
I've been working in the medical device field for over 25 years, including 8 years inside notified bodies.
00:00:32.880 --> 00:00:34.320
Let's get started.
00:00:44.399 --> 00:00:46.880
Let's not beat around the bush.
00:00:47.039 --> 00:01:01.119
The technical documentation submitted to the notified body must of course meet the requirements of annexes 2 and 3 of the MDR, but it must also be understandable for the product assessor who has to review it.
00:01:01.759 --> 00:01:06.719
Believe in someone me who has held this role in the past.
00:01:07.280 --> 00:01:24.400
After all, what do you do when you receive an incomprehensible gas or electricity bill and you try to find a specific piece of information like your monthly consumption, but you can't, because it's buried under a mountain of other data?
00:01:24.640 --> 00:01:26.400
You get annoyed, right?
00:01:27.359 --> 00:01:30.079
Don't annoy the product assessor.
00:01:30.239 --> 00:01:35.760
Not because they delicate souls, but because it will cost you time later.
00:01:36.079 --> 00:01:41.359
And with the MDD deadlines approaching fast, there's no time to waste.
00:01:41.840 --> 00:01:51.439
So put yourself in the assessor's shoes and evaluate your technical documentation from the perspective of evidence accessibility.
00:01:51.680 --> 00:02:03.439
We had a rule: if it takes more than 3 minutes to find something right and on conformity, it's very likely your product assessor will think the same way.
00:02:04.079 --> 00:02:23.520
If you need a solid reference on how to properly structure technical documentation full of practical tips, I recommend the Team MB guideline Best Practice Guidance for the Submission of Technical Documentation under Annex 2 and 3 of Medical Device Regulation.
00:02:23.759 --> 00:02:27.759
It's easy to find with a simple online search.
00:02:28.000 --> 00:02:41.759
The document was drafted by notified bodies, I personally took part in the founding working group to help manufacturers prepare a technical file that is organized, complete, and effective.
00:02:42.479 --> 00:02:57.360
Unsurprisingly, the guideline identifies the two most common cases of delays in assessments as first submitting incomplete documentation, second, lacking of a consistent structure.
00:02:57.599 --> 00:03:01.120
The information exists but is hard to locate.
00:03:02.400 --> 00:03:09.120
Avoid not searchable PDFs, for example, fully scanned documents.
00:03:09.439 --> 00:03:19.840
I remember a manufacturer submitting 1500 pages of non-searchable files with a table of contents that did not include page numbers.
00:03:20.080 --> 00:03:28.159
It was a nightmare, the kind that gives you carpal tunnel syndrome from scrolling just to find what you need.
00:03:28.479 --> 00:03:31.039
Pay close attention to completeness.
00:03:31.280 --> 00:03:37.520
If a supplier sends you a summary test report without raw data, expect a finding.
00:03:37.919 --> 00:03:41.520
Prevention is always better than cure.
00:03:41.680 --> 00:03:46.000
And don't be fooled by confidentiality excuses.
00:03:46.400 --> 00:03:58.800
The notified body is required by its own regulatory obligations to maintain confidentiality and integrity of all information received so your data is safe.
00:03:59.039 --> 00:04:05.599
When you revise the technical file, make it immediately clear what you changed and why.
00:04:05.840 --> 00:04:08.400
Make the assessor's job easier.
00:04:08.639 --> 00:04:16.639
Don't fall into the trap of thinking that making their work harder will make them more lenient.
00:04:16.959 --> 00:04:23.759
They'll simply issue more non-conformities because they can't understand what they are reading.
00:04:24.079 --> 00:04:36.399
Choose any system you like, underlining, sidebars, color coding, but highlight your changes and document this practice in your quality management system.
00:04:36.720 --> 00:04:44.560
When responding to findings, write as if the recipient were reading that report for the very first time.
00:04:44.959 --> 00:04:47.360
Don't take anything for granted.
00:04:47.680 --> 00:04:54.959
Better one repetition too many than one concept too few, because the burden of proof is up to you.
00:04:55.680 --> 00:05:05.279
When you send a technical documentation in response to findings, always send the entire technical file, not just the modified sections.
00:05:05.759 --> 00:05:17.839
Otherwise, you risk compromising the integrity of the documentation on the notified body's side, which is never a good thing for either party.
00:05:18.240 --> 00:05:26.879
If you need to translate the technical documentation, because the notified body requires a different language, do it carefully.
00:05:27.199 --> 00:05:34.560
No need to explain why, but don't rely on crude online translation tools.
00:05:34.800 --> 00:05:41.439
Otherwise, at best you'll waste time with back and forth questions and clarifications.
00:05:41.920 --> 00:05:44.800
Watch out for duplicated information.
00:05:45.040 --> 00:05:46.480
There's plenty of it.
00:05:46.720 --> 00:05:55.920
For example, the intended use and the basic UDIDI may appear multiple times in different sections of the technical file.
00:05:56.240 --> 00:05:57.839
Be consistent.
00:05:58.319 --> 00:06:03.279
Always double-check that these elements are reportedly identically.
00:06:03.519 --> 00:06:10.000
Use your WordPressor search or bookmark function to quickly jump to the relevant sections.
00:06:10.959 --> 00:06:19.759
One last tip carefully review the notified body's own regulations before submitting your documentation.
00:06:20.160 --> 00:06:24.639
They are always available on the notified body's website.
00:06:25.120 --> 00:06:34.959
These documents often contain additional requirements, such as language requirements, that are contractual and therefore mandatory.
00:06:35.120 --> 00:06:41.600
And don't forget, you can always request a structured dialogue with the notified body.
00:06:41.839 --> 00:06:53.600
This is not consultancy, which the notified body cannot provide, but clarification on regulatory procedures, which the notified body is obliged to give.
00:06:54.000 --> 00:07:08.399
It is provided for in the MDR itself, Annex 7.129, theoretically at no additional cost as stated in MDCG 2019 6.
00:07:11.759 --> 00:07:16.480
The important thing is to avoid asking how do I become compliant?
00:07:16.720 --> 00:07:21.199
Because that is consultancy and the notified body cannot answer.
00:07:22.000 --> 00:07:27.040
In short, help the product assessor to help you.
00:07:27.439 --> 00:07:34.560
Put them in the best possible position to do their job and everything will run smoothly.
00:07:39.839 --> 00:07:44.079
You've just listened to an episode of Compliance Espresso.
00:07:44.319 --> 00:07:47.839
The next episode will be out in two weeks.
00:07:48.079 --> 00:07:55.600
If you have comments or suggestions, feel free to write me at nicola at focci.it.
00:07:56.160 --> 00:07:57.759
Thank you for listening.