अनुलिपि 🔗
Pierre: let's go. Hi Megan.
Meg Provenzano: Hi, thank you for having me. This is great. I'm excited.
Pierre: Yeah, I'm very happy to have you. So in today's episode, we are going to talk about backtail endotoxin tests. And what I found strange about it is that on paper, when I read the protocol, it looks easy, pretty straightforward. But whenever I talk to people, people in the lab, each time they told me, it's not that easy. It's a bit tricky. The test seems to be very sensitive. There are lot of small details to pay attention to. And sometime when it goes wrong, the investigation can take a long time. So I'm very happy to have you with us today, Meg, because you have a lot of hands-on experience for performing BET testing as a lab technician and now as a product manager for a BET supplier. So I guess you're the perfect person to answer our questions. So welcome. And thank you very much for joining us today. to talk about the test itself, can you introduce yourself?
Meg Provenzano: Of course. Yeah, so I'm Meg Provenzano. I'm the product manager for our BET product line here at Veolia, for our endotoxin product line. I've been doing endotoxin testing or been around it for over 15 years, which kind of dates me. But yeah, I I started off doing quality control testing. so looking at in process materials, final product testing, a little bit of sterility as well. And then moved into technical support where I did a lot of kind of contract lab testing of many different sample types. and anywhere from doing like interference enhancement kind of screening to final product release or validations even. and then answering customer questions as well about their endotoxin testing. So troubleshooting, stuff like that. And then I moved into product management because you get to dabble a little bit in everything, which is fun and you get to kind of shape the user experience. but oddly enough, I started off as going to school for marine science. So I really wanted to be a dolphin trainer. So a little bit Little bit far away from where I started, but I love it. I love endotoxin testing. I love talking about the horseshoe crabs still. so yeah. I'm excited to be here.
Pierre: Okay, you're walking with dolphins, but much, much smaller.
Meg Provenzano: Right.
Pierre: Okay, so I'm sure you've got a lot of experience and a lot of tricks to share. And this is what we want to talk about today, those small tricks to make sure that the bacteria on the toxin are going well.
Meg Provenzano: Yeah.
Pierre: Maybe before we go further, can you explain us in simple words, what is an endotoxin?
Meg Provenzano: Yeah, so this one trips everybody up, right? Or people get the concept of endotoxin testing a little bit, but really endotoxin comes from the outer cell membrane of or outer cell wall of a gram-negative bacteria. The thing that kind of trips everybody up is that this is a biological assay. So there is a 50 to 200% window of whether or not your results are going to be correct.
Load full transcript
So, I think that's one thing, you know, you mentioned the complexities and it it looks clear on paper, but when you get into it, there's so much gray area for endotoxin testing. So, another thing that I like to mention too is that endotoxins are pyrogens, which means that causes a fever or it can cause even sepsis and death in patients. But not all pyrogens are endotoxin. Pyrogens can also come from gram-positive bacteria. So that's an important distinction where the gram-negative bacteria are the ones that really cause the problems and can cause death. That's why we want to make sure that we're measuring the amount of endotoxin in your injectables or medical devices or anything that really comes into contact with the blood or cerebral spinal fluid of a patient. And the whole reason why we do endotoxin testing is for patient safety. So we want to make sure we're keeping everybody's grandma and kids and even your dogs safe.
Pierre: OK, that's good. So this is
Meg Provenzano: Yeah.
Pierre: something we really don't want to have in our product.
Meg Provenzano: Right.
Pierre: But can you tell us more about how do we test it? mean, there are many different tests out there. But
Meg Provenzano: Mm-hmm.
Pierre: what's the principle? Because most of the people here are microbiologists. So we use agar plates. We put the sample. It grows. OK, we see the colonies. This is positive. I mean, it is something. But
Meg Provenzano: Mm-hmm.
Pierre: here, this is only small pieces of a. bacterial cells. So how do you detect them?
Meg Provenzano: Right. So originally the the way that we would detect was rabbits. So they would inject the drug product into the rabbit and then check to see if the rabbit spiked a fever. obviously that's not ideal because nobody wants to hurt bunnies. so in the nineteen seventies, Dr. Frederick and Dr. Bang discovered that the horseshoe crab blood actually gels in the presence of gram-negative bacteria. So over time they developed this test into the LAL test, the lumulus amoebicite lysate test. And the principal mechanism is that we use LAL and it it basically gels when there's endotoxin. Over the years, the LAL manufacturers have refined that into a kinetic assay where we're not just looking at the presence or absence of a gel. We're actually looking at the kinetic measurement over time of the of the LAL using a turbidometric kinetic turbidometric method or a kinetic chromogenic method. So we're looking at either the change in turbidity or the change in color over time. And this has evolved over many, many years too. So that's kind of the principal way that we look at it. I think L A L is the main one, but there are emerging technologies out there. there's microfluidics and then there's also recombinant technology, which is a synthetic version of the LAL.
Pierre: Okay, so many different tests are available
Meg Provenzano: Yeah.
Pierre: and as you described them, it looks still pretty easy. So what's the problem?
Meg Provenzano: I should
Pierre: What's for you? What are the main issues with the bacteria on the toxin test?
Meg Provenzano: boy. there are so many. Most everything interferes with this assay. So your drug
Pierre: Okay. Okay.
Meg Provenzano: product, your drug substance substance, your raw materials, any accessories that you might use in your in performing your test can interfere. So there's so many different types of interferences on this assay. There's metals. Metals can cause different kinds of Interference. Chelators can cause different interferences where you may need to use specific buffers to overcome that chelator effect. what else? Plastics. The biggest thing that I've actually seen over the years is not necessarily that the drug product or drug substance itself is the problem. It's usually the accessories. So polypropylene plastic, for example. Is widely known to cause interference with this assay. Unfortunately, people still will choose accessories that are made out of polypropylene plastic. I've seen serological plastic pipettes that are used a lot that cause interference. They can either leach out endotoxin or they can add kind of particles that interfere with the assay. Adding bubbles when you're pipetting also can cause interference. So you want to be very careful when you're pipetting. one common thing with kinetic assays is when you're adding either your lysate or your samples to the plate, you can get splash out. So then you get contamination into your other samples or causes CV issues, PPC issues, all kinds of stuff. So it's just a very complicated test where you have to be very cognizant of of what you're using, what's in your product as well. because it different incipients can cause the different inferen interferences as well.
Pierre: OK, so interferences can be caused by the material that you are using outside of the product sample, but as well, it can come from the product. So interferences
Meg Provenzano: Right.
Pierre: are everywhere. But what exactly is interferences? What does it mean? That mean you have false positive results? Or what does it mean for the results?
Meg Provenzano: great question. So interference shows up typically in your positive product control. so what you'll see is if it's a very inhibitory material or sample, you'll see PPC recovery below 50%. If you have an something that's enhancing in your test material or sample, it'll show up as either near 200% PPC recovery or greater than 200%. recovery. So that just means that these things are kind of getting in the way of the LAL enzymatic reaction and either enhancing that reaction or inhibiting it. And that can show up as false positives. We see that a lot with beta glucan interference. So if there is cellulose materials either in your pipette tip or in some Part of your process where there's filtration or something like that, that will definitely show up with beta glucans and cause a false positive and cause enhancement. So there's a number of different ways that it shows up
and that you can detect this kind of interference. But typically we look at the PPC recovery. Sometimes it can even show up in your coefficient of variation percent as well. So kind of looking at the C V percent between each sample or the s the PPC samples. we see that sometimes too, where it exceeds the limit and then you kind of have to figure out what's going on there and it can be parti particulates or or something else going on.
Pierre: and what would be your best advice for the users to make sure that they do not have that much interferences.
Meg Provenzano: my best advice. first of all, never ever ever use polypropylene plastic. Ever. second advice would be to make sure that you're testing throughout the manufacturing process for endotoxin, but also, I mean, dilution in LEL reagent water is kind of your best bet. It's usually the easiest way to overcome interference where we want our drug products and drug substances to behave more like water because we know that we can test water with LAL and there's no typically no issues. but dilution in water and then also using endotoxin free materials and verified endotoxin free materials in your testing and sampling process. That's very important because Endotoxin free or pyrogen free doesn't always necessarily mean that it's qualified for the level of endotoxin testing that you're doing.
Pierre: So you mean that even for consumable, where it is mentioned endotoxin free, you can still have endotoxin on it? OK.
Meg Provenzano: Yes, absolutely. So one great example that I have is going back to serological plastic pipettes. Typically they say that they're endotoxin free, but then if you kind of drill down into the certificate, it'll say less than point zero six E per mil. and if you look into it further, they may be even tested with T A L, which is the tachypleis amoebocyte lysate that comes from the APAC region and those horseshoe crabs are a little bit different than the limulus horseshoe crabs so they behave a little differently. So if you're testing a kinetic assay for example down to 0.005 and you're using plastic pipettes that are rated to 0.06 those aren't really you know good enough for your assay so you have to really look into the materials and how they're tested. but also I always prefer to use the glass, serological glass pipettes anyway, because those can be deepyrogenated where you can't deepyrogenate plastic, otherwise it'll just melt.
Pierre: Okay, that's a good advice.
Meg Provenzano: Mm-hmm.
Pierre: I've been told that adsorption can be an
Meg Provenzano: Yeah.
Pierre: issue as well for endotoxin. What is it exactly?
Meg Provenzano: Yeah, that actually comes from the polypropylene plastic. And it can come from other materials as well. But that just means that the endotoxin that may be present in your sample is getting kind of trying to figure out a definition without using the word, right? But kind of being brought into the material. So you're losing endotoxin in your sample, because
it's getting kind of going back into the the tube, for example, your polypropylene tube, where you might get false negatives in that situation. So you really want to be careful about what dilution tubes you're using, making sure that they're polystyrene or borosilicate glass, because the glass, the type of glass matters as well in this assay.
Pierre: can you check that you had absorption or not? Is there any test you can do quickly to make sure that the consumables you're using are not absorbing the endotoxins?
Meg Provenzano: Yeah. I would recommend doing if you're if you suspect that, do a spiking study. So this can be something really simple where you add a known amount of endotoxin, either from your RSE or CSE that's matched with your LAL, and then wait over time, you know, typically one hour, and then measure the amount of endotoxin in that. Like if you're looking at Let's use the serological plastic pipettes again as an example. for those, I would soak them in LEL reagent water and add a known amount of endotoxin to that soaking or to that ex extract, and then test the extract after one hour to see if what I put in there comes out. So that I'm I'm making sure that the spike is recoverable and not being adsorbed into that plastic.
Pierre: Okay.
Meg Provenzano: If you're looking at dilution tubes, you could do something similar, but or you could just add, a known amount of RSC into the dilution tube and see if you can recover it after one hour as well.
Pierre: OK, so that's a quick and easy test to run. Each time that you change tubes or you change your pipette, this is something that should be done just to make sure that you do not have this adsorption issue. Any other advice? Yeah?
Meg Provenzano: Right. Right. I would also Yeah. good segue. I would also recommend if you do change pipette tips or any other materials, even if the manufacturer changes something in one of your excipients, to do a test on that as well. So make sure, like, for example, the excipient, if there's a change in that, that nothing interferes with the assay or that you're not introducing any endotoxin into the manufacturing process. But if you change pipette tips or dilution tubes or something like that, then you can do an easy extraction test. like pipette tips, I would take a random sampling of like 10 pipette tips, soak them in LRW, and then test that LRW at the end of one hour. and just make sure there's no detectable endotoxin. funny story is we tested For customer, some serological plastic pipettes. Again, these are the bane of my existence. so we we ended up doing an extraction, and I think it was like six out of ten of the plastic pipettes had endotoxin above 0.06. So they weren't really, you know, acceptable for the assay. But using that extraction method is really helpful. And then in the end you're just testing water, so it's pretty pretty simple.
Pierre: OK. Is there any tricks on the dilution as well? Because the volumes that we are using are very small, is there anything we can do on that?
Meg Provenzano: For for me, I always recommend using milliliters over microliters, just because if you make a mistake with a millilit with a microliter, let's say, that's a huge mistake compared to a milliliter. So I like to recommend that where possible, I know there's, you know, limitations in some some sample types, but where where it's possible, use milliliters rather than microliters. So say like one milliliter into nine milliliters for a one to ten dilution or 0.5 even into 4.5 milliliters. it'll give you a lot more wiggle room to make a mistake because we're all human. I personally like to use serological glass pipettes to make my dilutions. I find like that you know you can kind of get the meniscus where you typically get it, so as long as you're consistent with it, it'll work. I have a lot of colleagues that like to use pipette tips, but I always recommend using the serological glass just because that's kind of the gold standard and you're not introducing or allowing the sample to come into contact with plastics as much as possible. borosilicate glass dilution tubes are also kind of the best if you can if you can get there. and then another thing is not all pipette tips are created equal. I've seen some that cause a lot more interference than others. So there is, some pipette tip brands that are ideally suited for endotoxin testing. So I always recommend to to look at that as well.
Pierre: OK. OK, that's a lot of advice.
Meg Provenzano: Yeah, West three.
Pierre: Have you seen any other common mistake that we can do with the bacteria on the toxin tests?
Meg Provenzano: Some other common mistakes. let's see yeah, here's a good one. in the microbiology lab, there's usually pipette tips that require filters because when you're streaking plates and stuff like that, you wanna make sure that you're protecting your pipetteders. But those pipette tips or things that are suited for other microbiology. tests are not necessarily suitable for endotoxin testing. So you want to make sure that everything is free of of filters, including your your serological pipettes, your pipette tips, stuff like that, where you're really making sure that you're kind of separating out the endotoxin testing materials from your other microbiology tests to make sure that that you're using the appropriate materials for each Assay.
Pierre: OK, because in the filter you can get some interferences from the filter itself?
Meg Provenzano: Right. Even if the sample doesn't come into contact with the filter, or even if your standards don't, it can still introduce beta glucans to your sample or your standards or or whatever you're pipetting. So you really want to avoid filter filters in those. Another
Pierre: Okay.
Meg Provenzano: common thing that I've seen as well is using laboratory water instead of LAL reagent water. so like you know the the laboratory system to to kind of filter the water, make it like RO water or something like that,
that's not really suitable for the endotoxin test. We want to make sure that we're using LAL reagent water from the same LAL manufacturer also. I've seen some some issues with that in the past where people use whatever they have laying around for LAL reagent water, with their LAL, but really you should kind of go to the LAL manufacturer and use their their buffers, their reagent water, everything kind of together just to make sure that you're not adding anything in there because the LAL manufacturers do each have their own formulations. and they kind of they have their own, you know, own type of water as well. So the the buffering capacity of the LAL of the water could be a little bit different from manufacturer to manufacturer.
Pierre: Okay, this water is used for the dilution.
Meg Provenzano: Right. Either dilution or rehydrating the LAL. So you wanna
Pierre: Okay.
Meg Provenzano: make sure, you know, that lab laboratory water may not really go with the LAL. Or it could cause some interferences
Pierre: Okay.
Meg Provenzano: in your dilutions as well.
Pierre: OK. So now I've got a question from back when you were into the lab. What
Meg Provenzano: I guess.
Pierre: was the most difficult matrix that you had to test, and how did you make it visible?
Meg Provenzano: this one this one brings me back. there was a sample and it was like a sur a surgical kind of I think it was made out of pig skin. I'm not a hundred percent sure, but they used it for wound closures and surgeries. So obviously it has to be endotoxin free, but there was a lot of interferences from this. So it had a really low endotoxin limit, which meant that we had to extract it in a really small volume of water. And I want to say this took us, you know, me and a few others in the lab a couple, two to three weeks, I think, to figure out how to test this, with being able to stay below the the maximum valid dilution in the endotoxin limit. So ultimately. what we ended up doing was taking the lid of a ninety-six well microplate and putting the the pig skin or the wound closure whatever inside of the ninety-six well microplate lid and then we added LR up LRW. We let it soak for an hour, took the extract off. I think we even needed to use like a dispersing agent and a glucane blocking buffer on this one, but it took so many different tries and I felt like I was banging my head against the wall trying to get the non-interfering dilution of this of this thing. but ultimately we figured it out and were able to pass on the results in the method to the customer. So it was a a long road and I think I still have a little bit of PTSD from that one.
Pierre: It seems that in the BET test, one of the most difficult part is the sample preparation as well. So this is where most of the difficulties, mistakes can happen as well.
Meg Provenzano: Mm-hmm.
Pierre: What do we think about that?
Meg Provenzano: Yeah, absolutely. I mean there's you know, proteins need heat denaturization, usually heat treatment. So there's just a long a lot of different ways to overcome interference. There's a lot of typical ways too. So like if you know that you have are testing a protein, you know that you need to heat treat it at however much sixty, seventy degrees C. maybe even 80 degrees C. but it's not the same for every sample. So you really have to figure out the sample preparation for each individual sample. Another thing, another piece of advice I like to give is that not every manufacturer is the same either. So if if you're making saline, you may be making saline a little bit differently than I make saline, right? And that's just an example. But so we want to make sure even between different types of products that you're kind of going through in looking at the inhibition enhancement profile of that sample. So typically what I would do is do like tenfold dilutions to start off, unless I know the MVD or endotoxin limit and just go from there and see what I can do and then throw in different buffers. Look at the pH of the sample mixed with the LAL. stuff like that. Yeah.
Pierre: Okay, so there are many ways of getting rid of that interference and making sure it works. I
Meg Provenzano: Right.
Pierre: guess it's just a question of experience.
Meg Provenzano: Exactly. And experimentation, honestly. You know, sometimes you just have to play around with it and figure out what the best best method is. USP chapter eight ten eighty five has some good advice in it as well, looking at different materials, but again, it's more you just have to play around with it and figure out what works in your situation with in your laboratory even.
Pierre: OK. So at the beginning of the podcast, you mentioned that you work now for a BET supplier. Can
Meg Provenzano: Right.
Pierre: you just tell us more about what type of solution do you have for BET?
Meg Provenzano: Yeah, we have a pretty novel solution. we use microfluidic technology in centripetal force to basically evenly distribute LAL in sample to each other in a one-to-one ratio. So we have a microfluidic disk that has twenty-one samples and five a five-point standard curve. it'll run the five up to a five-point standard curve, I should say. So you can either run a fifty to point zero zero five or a five to point zero zero five or however you want to configure that. And then we have and it runs the standard curve in triplicate with negative controls each two. so that's dried and predeposited on the the disk. And then the PPCs are pre-deposited on the disk as well. So a lot of problems that come about with the BET assay are with your standard curve. So human error is a big problem. We haven't touched on that yet, but I might make my standard curve a little bit differently than you do. So your reaction times might come out a little bit differently. So what we were trying to do is really minimize the human error aspect as much as possible with this microfluidic disk. So it's called the Sievers Eclipse. and we have our own
instrument, really great software that's very intuitive, and then the microfluidic disc. so like I said, you can run up to twenty one samples on it and you don't have to, but if you want to you can. so yeah, it's super
Pierre: Okay, so.
Meg Provenzano: easy to use too.
Pierre: I've been told that you do not sell the reagents, the BET reagents. So you only sell, I would say, the disk and the equipment
Meg Provenzano: Yeah.
Pierre: to read the disk. So what type of reagent can the user use?
Meg Provenzano: yeah, it we are reagent agnostic, we like to say. So you can use any kinetic chromogenic reagent out there, commercially available. We're also compatible with all commercially available recombinant cascade reagents. And then we do work with one kinetic turbidometric reagent. So the other nice benefit of the eclipse system is that we only use one milliliter of LAL. So if you're looking at sustainability goals and stuff like that, and reducing the impact on animals, then you can reduce that by ninety percent with the Sievers Eclipse microfluidic disc, which is really cool.
Pierre: What volume can you say it again?
Meg Provenzano: One milliliter for twenty-one samples in a five point standard curve. And if you
Pierre: Okay. Okay.
Meg Provenzano: compare it to a ninety-six well microplate, that uses ten milliliters for the equivalent.
Pierre: OK, good to know.
Meg Provenzano: Yeah.
Pierre: Now, mean, you are the supplier, so of course you have many advantages. But for you, what are the main two or three advantages of your solution?
Meg Provenzano: I'll tell you what, I've been I was actually in the lab recently doing ninety-six well microplates, and I think the biggest advantage is not having to create your own standard curve each day. that can take thirty to forty five minutes because you have to vortex your C S E or R S E for so long and then make your dilutions. The other big advantage that I I find is the pre-deposited PPCs. So with LAL, you typically will hot spike in the microplate with 10 microliters. With RCRs, you spike with five microliters. And I don't know about you, but I'm getting old and my eyesight's not that great anymore. So being able to tell whether or not I spiked with 10 microliters or five microliters is really difficult. So back in the day I could I could kind of see the volume change and stuff like that, but now it's I you know it's really difficult to see. So having the PPCs dried on there is really beneficial. The other thing is that for each sample you only have to pipette one time. So we use 55 microliters per sample. And that gets equally distributed into four different optical wells, two for your sample, two for your PPCs. And I only have to pipette once. So going from the 96 well microplate, or even gel clot, it's so much nicer to just be able to do that one pipetting step. and same with the LAL. You only have to pipette one milliliter into one port, and then it gets equally distributed. across the entire plate. So that's really nice, especially after having done all kinds of B E T tests, by the way. I've done gel clot, kinetic, everything.
Pierre: OK, sounds good. And
Meg Provenzano: Mm-hmm.
Pierre: now I've got a question you might not like, but what is the limitation of this technology? If you had to mention one.
Meg Provenzano: I I know I'm the supplier, but I guess the biggest limitation is some some labs may not have twenty one samples at a time. they may have, you know, ten or eight or something like that. So in the end we recommend batching your samples. So maybe holding on to samples and running them all at once at a later date.
Pierre: So you keep the sample for one or two days, and whenever you have 21 samples, you can run them.
Meg Provenzano: Right.
Pierre: That holding time, is it OK? Or is there any chance that the undertoxin fly away? I don't know if they fly away, but they disappear.
Meg Provenzano: Yeah, in that case if you do have to like batch samples and really we see an advantage advantage of using the disc at if you have 12 samples even. So we would recommend doing a whole whole time study. So you might want to just hold the samples in those containers and then test at various time points, right? So if you're gonna be holding, and this is true honestly for any process. Anytime that you have to hold your samples in a container, you want to make sure you do a whole time study. So testing it at however many time points you deem necessary for the time period that those samples are going to be kept in that container. So if you're batching samples or if you're holding your samples for seven days, then you might want to test at zero, three, five, seven days. It maybe even a worst case scenario of maybe you're sick and and you can't get into the lab to test maybe nine days, right? So, I it's typically pretty easy.
Pierre: So basically you spike your sample, you keep it into the container for X days, and then you test it on a regular basis to make sure that you are able to recover them. That's
Meg Provenzano: Right.
Pierre: the way it is done. Sounds easy.
Meg Provenzano: Yeah, you can do Yeah. It usually is pretty easy. you could do two things actually. So you could spike and then you could also simultaneously have just the samples in the containers so that you're looking at any container effects kind of as a control, but then also spiking to make sure that you can recover that endotoxin as well.
Pierre: OK, OK, OK, makes sense. In the endotoxin world, now we hear more and more about the recombinants.
Meg Provenzano: Mm-hmm.
Pierre: Can you just, you already mentioned it a little bit, but can you, even though you do not sell any reagents, so, but can you explain us how it works and why it's different from the LAL?
Meg Provenzano: Yeah. Sure, sure. There's two different types of recombinants out there, actually. there's the recombinant factor C. so if you look at the enzymatic cascade, there's multiple different factors for the LAL. There's
recombinant factor C, factor B, there's even a factor G, and then a coagulin. So recombinant factor C's only have that one part of the enzymatic cascade. So because of that. That's a fluorescence assay, which requires different kind of preparation of that reagent. you also have to look at the gains for the fluorescence reader for the RFC and adjust it kind of each time you get a new lot. And then there's recombinant cascade reagents, which mimic the enzymatic cascade of the limulus amoebocyte lysate, where RFC is typically from the carcinoscorpius, or or maybe from a different type of horseshoe crab. So RCRs, recombinant cascade reagents, have recombinant proteins for each of the different factors in the enzymatic cascade of the LAL. So they are a chromogenic method, which means that they're they're a little bit more similar to LAL. We see a little bit less lot to lot variability with the recombinant cascade reagents than we do either with, you know, LAL, TAL, or even sometimes RFC. So there's some benefits you know, on both sides. And one thing, you know, we had an endotoxin master class a couple weeks ago in San Diego, and one thing that kind of came up from one of our speakers was, you really want to look at each reagent for your sample type. Don't just pick one reagent to kind of broadly paint across all samples because every sample behaves a little bit differently. You might need a turbidometric reagent over a chromogenic reagent, for example. So I just wanted to mention that because I think a lot of the
Pierre: Yeah.
Meg Provenzano: times to people think, I'm just gonna go with kinetic chromogenic. But, there's different benefits to each reagent type.
Pierre: OK, so one-feet-all does not really exist in endotoxin, and the
Meg Provenzano: Right.
Pierre: solution might be in using different types of solutions. Even I've heard that from a user. That was a contract lab, so they were using different types of technologies. And even the same type of technology coming from two different suppliers, they told us that they have different types of results. One was compatible with the product and not the other one, and they could not explain why. But it was working
Meg Provenzano: Yeah.
Pierre: with one and not with the other one. Okay.
Meg Provenzano: Yeah. And like I said, each manufacturer has their own proprietary formulation and that is for recombinants and for LAL. So you just have to kind of again play around with it and see what works best for your method and your products.
Pierre: OK, that's a good advice, even though it's not that easy for the lab to have different type of technology, different suppliers. I'm sure the suppliers,
Meg Provenzano: Yeah.
Pierre: love it. But for the lab, it's not always easy to deal with different technology suppliers because the volume might be different. You need to train the people as well. OK.
Meg Provenzano: Right. Yep. That's true.
Pierre: In the future, we might see more and more recombinant solution being
used in the lab. So we are slowly transitioning from LAL to recombinant. The customer that you are following, which are probably doing that transition, what is their strategy to go from one to the other one?
Meg Provenzano: Yeah, we've we've had a few customers actually that that looked at this. So one strateg you know a couple of customers have tested multiple recombinants alongside their LAL to make sure that they're they are choosing the kind of the best one for their process. but then also, you know, looking at the vendor's primary validation package is really important. And seeing if there's any gaps. So if you typically see B sapasia in your manufacturing process and the manufacturer didn't test B sapasia, for example, with their RCR, then it's always a good idea to close that gap and make sure that anything that you typically see that, you know, when you do unfortunately get a contamination event, that you can detect that contaminant. Because again, the whole goal of B E T is patient safety. So we want to make sure that unfortunately when we do get contamination, it happens that we can detect that contaminant. And every single manufacturer has fully different you know, morphologies of the the bugs too, because we're constantly trying to kill them. So they're trying to to evolve and be able to protect themselves.
Pierre: OK, one of the strategies I see is that for the new product, of course, they validate the new product with the recombinant one. They do the water as well first. And then
Meg Provenzano: Mm-hmm.
Pierre: little by little, they gain confidence with product with water. And they will slowly transition all the old product to the recombinant. I mean, that's a long way that might take 10, 15 years. But they are going slowly. toward the recombinant because what we've been told, I don't know if it's the same case for you, is that at one point, same as for the rabbits, they will probably, we won't be able to use the horseshoe crab blood anymore and we will have to use the recombinant. So at one point we will all have to use recombinant. It's just a matter of how long. do we have to make sure that we slowly transition from one solution to the other one?
Meg Provenzano: I honestly think that LAL is going to be used for many more years. I don't I think it's gonna
Pierre: Okay.
Meg Provenzano: be a really long transition. And then another
Pierre: Okay.
Meg Provenzano: thing to to keep in mind is that the biomedical industry protects the horseshoe crabs. So they actually have put into place laws in certain areas that where the horseshoe crab is used for biomedical bleeding, for example, to get the LAL, that you can't use those horseshoe crabs for baiting. So the biomedical industry takes really good care of the horseshoe crabs. So if we completely move away from LAL, then there those protections are no longer going to be there. so we want to make sure that, you know, that we are taking care of this precious resource because these crabs have evolved for over four hundred and fifty million years and been around. So I think yeah people are going to be moving to recombinant, but I think I've also heard that some products don't work with recombinant. So you have still have to use LAL.
Pierre: Okay.
Meg Provenzano: or vice versa, you know, again it goes to picking the best reagent for your product and process.
Pierre: So as you mentioned earlier, we might end up with different technologies testing different type of product in the same lab. Okay.
Meg Provenzano: Right. Yeah, I think so.
Pierre: Okay. So we'll see. We'll see in 10 years from now, we can do another podcast and see where we are.
Meg Provenzano: That sounds good. Let's do it.
Pierre: Okay, Megan, so thank you very much for answering all those questions. It was very clear. I hope it will be very clear for everybody, but you brought many pieces of advice that could change the way we do endotoxin tests into the lab. So thank you very much for that.
Meg Provenzano: You're welcome, definitely. And if anybody has any questions, feel free to look me up on LinkedIn or reach out. I'm always happy to answer endotoxin questions. Or if you want to hear my soapbox on horseshoe crabs even further, please reach out to me.