00:00:00.080 --> 00:00:04.959
Welcome back to Measure Matters, the podcast where we dive into all things metrology and calibration.
00:00:05.200 --> 00:00:08.080
I'm Sally Pittman, the content specialist here at MENSER.
00:00:08.240 --> 00:00:13.519
And today we're unpacking the differences between two essential classes of calibration, field and lab.
00:00:14.080 --> 00:00:18.879
Today I'm joined by Marshall Andrews, one of our sales application specialists with a wealth of product knowledge.
00:00:19.199 --> 00:00:28.399
He's here to shed some light on the key distinctions between field and lab calibration, the circumstances that can influence which route you choose, and even highlight an exciting new product.
00:00:28.719 --> 00:00:31.359
It's a topic that impacts a lot of metrology professionals.
00:00:31.440 --> 00:00:32.799
So let's jump right in.
00:00:33.119 --> 00:00:37.039
You're listening to Measure Matters, a podcast by Mencer.
00:00:38.640 --> 00:00:39.600
All right, Marshall.
00:00:39.679 --> 00:00:43.920
So before we get started, um I happen to know that you're a bobcat from Texas State.
00:00:44.079 --> 00:00:45.039
Is that is that true?
00:00:45.280 --> 00:00:45.840
That is true.
00:00:46.000 --> 00:00:51.359
I am a Texas State Bobcat, graduated way back in the day, age of the dinosaurs, I guess, uh in 2019.
00:00:51.759 --> 00:00:52.479
Age of the dinosaurs.
00:00:52.799 --> 00:00:52.960
Right.
00:00:53.039 --> 00:00:57.759
I got my degree in manufacturing engineering from Texas State back then, and then um joined the Navy afterwards.
00:00:57.840 --> 00:01:00.799
And after a short stint there, uh came here at Menzer.
00:01:00.960 --> 00:01:03.039
So now I'm an application specialist here at the company.
00:01:03.280 --> 00:01:07.840
So how did you go from the Navy into such a niche industry?
00:01:08.400 --> 00:01:09.680
Um, yeah, great question.
00:01:09.760 --> 00:01:12.239
So it's actually it's not what you know to you know.
00:01:12.319 --> 00:01:17.120
Um so I've always knew I wanted to get back into um my engineering background.
00:01:17.200 --> 00:01:25.359
Um so finding some sort of company where I could blend um that skill and trying to find a job too, where I love talking with people, I love the soft skills.
00:01:25.439 --> 00:01:31.680
So finding a way that I could blend uh my technical expertise in engineering and um kind of a sales background.
00:01:31.760 --> 00:01:34.000
I grew up somewhat of a sales family for my dad.
00:01:34.159 --> 00:01:36.400
Um, and then I also knew I wanted to be in the hill country.
00:01:36.480 --> 00:01:38.959
So trying to find a way to blend all these different pieces together.
00:01:39.040 --> 00:01:44.319
And luckily, Menzer was up on that list and you know, talking about it's not what you know, it's who you know.
00:01:44.480 --> 00:01:50.560
Um I had um some family friends through the company that was uh were able to get a good referral for me.
00:01:50.719 --> 00:01:55.280
Load full transcript
And um, so it's really important to network and and meet great people in life that can help push you forward in your career.
00:01:55.439 --> 00:01:56.239
Yeah, that's awesome.
00:01:56.400 --> 00:01:59.519
Sounds like you've got all the things you were looking for, hopefully.
00:02:00.000 --> 00:02:09.360
Um, so for anyone who's kind of new to the conversation, what exactly is calibration and why is it such a critical process across so many different industries?
00:02:09.599 --> 00:02:10.479
Yeah, great question.
00:02:10.639 --> 00:02:18.960
So um essentially calibration is um comparing an instrument or device under test to a reference standard or a primary standard.
00:02:19.039 --> 00:02:24.479
Um, you know, your goal there is to verify the accuracy and document any errors.
00:02:24.560 --> 00:02:31.520
And then occasionally what you need to do if the device under test is out of spec is you um need to correct it.
00:02:31.759 --> 00:02:42.879
And so um doing these things, then the importance behind it is um for things like safety, uh quality in your products, um, compliance, and um process efficiency.
00:02:43.039 --> 00:02:57.439
So um bad measurements um due to incorrect readings can lead to you know safety concerns in the sense that maybe uh an instrument's reading um one PSI, and this might be an extreme case, but maybe it's reading one PSI and it's actually experiencing 100 um pounds on it.
00:02:57.520 --> 00:02:59.439
That's again a very extreme case.
00:02:59.599 --> 00:03:08.000
Um, but you know, if a technician goes up and tests that and doesn't realize there's much more pressure on it than expected, it could lead to injuries.
00:03:08.240 --> 00:03:17.120
Um, you know, with product quality, a bad reading can lead to an entire batch of um whatever's being manufactured and having to be discarded and the loss of revenue there.
00:03:17.280 --> 00:03:23.680
Um, so a good simple phrase I like to think about is you know, calibration gives you the confidence that the instrument is telling the truth.
00:03:24.000 --> 00:03:24.479
Gotcha.
00:03:24.560 --> 00:03:25.280
That makes a lot of sense.
00:03:25.360 --> 00:03:30.960
So it's all about knowing that your instruments are really reading what is the actual pressure or whatever it might be.
00:03:31.120 --> 00:03:32.080
Yeah, that makes sense.
00:03:32.719 --> 00:03:40.000
Um so before we jump into strategy, can you walk us through some of the fundamental differences between field calibration and lab calibration?
00:03:40.319 --> 00:03:48.240
Yeah, so um I'd say that the big difference between the two, the main difference would be um your environment and the purpose.
00:03:48.400 --> 00:03:52.080
So, like in lab calibration, it's done in an extremely controlled environment.
00:03:52.240 --> 00:03:58.000
And here you're focusing on the highest um accuracy, low uncertainty, um, and repeatability.
00:03:58.080 --> 00:04:02.159
And so um this is again better for tighter uncertainty requirements.
00:04:02.319 --> 00:04:12.960
Whereas in field calibration, you're looking for um you're taking that calibration um device and going uh to the field where this instrument is installed rather than removing it.
00:04:13.120 --> 00:04:18.240
And so this is focused focused more on like convenience and downtime efficiency, et cetera.
00:04:18.480 --> 00:04:24.319
Um and so that's a lot better for your operational maintenance efficiency, having to reduce those effects.
00:04:24.480 --> 00:04:31.279
So um a good way to think about it is you know, your lab environments more for your control and your field environments more for um your practicality.
00:04:31.600 --> 00:04:32.000
Gotcha.
00:04:32.079 --> 00:04:32.879
That makes a lot of sense.
00:04:32.959 --> 00:04:35.439
So it kind of comes down to like convenience environment.
00:04:35.519 --> 00:04:38.000
And what was the third the third point you said again?
00:04:38.240 --> 00:04:43.759
Uh convenience environment, um, and the the purpose, like what are you doing with this?
00:04:43.839 --> 00:04:46.720
Um, how are you um using your device under test?
00:04:47.040 --> 00:04:47.360
Gotcha.
00:04:47.439 --> 00:04:48.800
Okay, it makes a lot of sense.
00:04:48.959 --> 00:04:53.279
Um, so lab calibration is treated a lot of the times as like the gold standard.
00:04:53.439 --> 00:05:02.000
Um so from your perspective, what makes the lab environment so uniquely capable of delivering that gold standard as opposed to field?
00:05:02.480 --> 00:05:03.519
Uh, another great question.
00:05:03.680 --> 00:05:07.040
So the lab kind of touching on what I just talked about.
00:05:07.120 --> 00:05:15.680
So the control over the environment itself, the control over the invariables, um, being uh temperature within the lab, um, cleanliness.
00:05:15.839 --> 00:05:20.399
So you don't want to um take a dirty sensor, come in and calibrate it.
00:05:20.480 --> 00:05:24.959
These um instruments like our 6050 we have here on the desk are very sensitive to contaminants.
00:05:25.040 --> 00:05:31.040
Uh so making sure that not only the lab is clean, uh clean, but also the devices you're you're calibrating are clean.
00:05:31.199 --> 00:05:32.399
Um, vibration.
00:05:32.560 --> 00:05:45.360
So a cool fact about our cow lab here at Minster, when I was um green, um learning all the different things around here, um, I found out that there's actually a section of our cow lab that the foundation is cut out from the rest of the foundation of the building.
00:05:45.519 --> 00:05:53.920
Um, and that's because um you don't want vibration from either us walking around, the the manufacturing operations going on the rest of the building.
00:05:54.000 --> 00:05:56.480
You know, we live here in San Marcos, so you know all about the trains.
00:05:56.560 --> 00:06:05.920
Uh, you know, we got a train track about probably half a mile up the road from here, but even vibrations from the train um that far away can affect these
calibrations can show up in the reading.
00:06:06.079 --> 00:06:11.120
So um considering vibration um in there as well, uh as well as setup consistency.
00:06:11.199 --> 00:06:25.360
So the importance of the orientation of device you're testing, you know, maybe it's vertical, horizontal, that can um have effects, as well as the the height of your calibrator versus your um the device under test, the head height is what we call it that all can also affect reading.
00:06:25.439 --> 00:06:28.079
And then the last piece, you know, um pressure stability.
00:06:28.240 --> 00:06:34.240
So another cool thing, I'm not sure if you've seen our Cal Lab yet, but actually to enter it, you got to go through two sliding glass doors.
00:06:34.399 --> 00:06:40.560
Um, and the reason being, so you walk in one door, you have to wait for that first door to close before it allows you to walk in the second.
00:06:40.639 --> 00:06:44.959
And the reason being because there's a pressure gradient between the two sides, whether it's high or low on one side.
00:06:45.040 --> 00:06:54.240
Um, but if they both open up at the same time, the pressure can uh fluctuate drastically, and that can absolutely show up in readings, especially when we're dealing with low pressure calibrations.
00:06:54.720 --> 00:07:02.560
Um, so fewer outside variables um allow us to get a much better um accuracy and repeatability when doing these calibrations.
00:07:02.879 --> 00:07:05.759
Um so the coist is the best choice.
00:07:05.839 --> 00:07:10.480
Uh lab calibrations are the best choice when we're considering um um higher uncertainty.
00:07:10.560 --> 00:07:12.319
So when uncertainty matters the most.
00:07:12.560 --> 00:07:18.720
Um and also when we need to make sure we have a good documentation and these calibration certificates.
00:07:18.879 --> 00:07:28.480
Uh so a good phrase I like to think about is that you know, in metrology, um, control equals confidence in what our um DUT is re-enarrangement is going to read.
00:07:28.639 --> 00:07:33.279
Um so the lab allows a product like our 6050 to perform at its best.
00:07:33.519 --> 00:07:33.759
Okay.
00:07:33.839 --> 00:07:36.720
And that confidence really does come down to repeatability.
00:07:36.879 --> 00:07:37.920
Is that is that right?
00:07:38.079 --> 00:07:39.279
And that controlled environment helps.
00:07:40.319 --> 00:07:51.759
Repeatability and also um lowering the uncertainty because if we've out all these different variables that are fluctuating the pressure, it's it's hard for us to be um be certain on what the the true readings are.
00:07:51.920 --> 00:07:58.319
So so again, uh lower uncertainty, the high accuracy, and then the being able to repeat those data points again and again accurately.
00:07:58.639 --> 00:07:59.920
Okay, that makes a lot of sense.
00:08:00.079 --> 00:08:10.800
Um, so then on the flip side, for field calibration, what do you think are the most compelling factors that might influence people trying to, you know, go on site for their calibrations?
00:08:11.120 --> 00:08:14.399
Uh the biggest piece for field calibration is going to be your downtime.
00:08:14.639 --> 00:08:29.120
Um, you know, um, for all these processes that require the instrument to be involved, if I have to remove that instrument from, you know, your your gas line or or your skid or whatever it may be, um removing that shuts down that line.
00:08:29.360 --> 00:08:33.519
And um, you have to remove the instrument and you have to ship it out to a caliber.
00:08:33.679 --> 00:08:35.679
Maybe you're fortunate enough to have a calibr on site.
00:08:35.840 --> 00:08:41.120
Uh, you have to wait on the turnaround time for that to be calibrated and then you know reinstall it afterwards.
00:08:41.360 --> 00:08:46.559
Um, so it's a ton of downtime, ton of potential uh loss and revenue there, um, et cetera.
00:08:46.720 --> 00:09:02.240
And so being able to take a field caliber calibrator like this, taking it directly to the instrument where it's at, um, running your calibration verification, troubleshoot, whatever it may be, um allows for your operations and uh maintenance teams to handle these um situations faster.
00:09:02.320 --> 00:09:12.799
And also um instead of having to um worry about the downtime of the line, all those other factors, logistics, et cetera, that I just pointed out, um, allows you to lower the total cost for these calibrations.
00:09:12.960 --> 00:09:25.600
Um, and then you know, field calibrator like this, not only for calibrating, as I just kind of mentioned, too, it's extremely useful for troubleshooting, um, uh routine verification, different maintenance checks on these devices.
00:09:25.679 --> 00:09:38.960
Um, and so like a good phrase uh that I like to say for these is that um field calibration brings the the capability of a um of a calibration like the 6050 to the field, to the instrument in its place where it's being used.
00:09:39.200 --> 00:09:40.559
Okay, that makes a lot of sense.
00:09:40.720 --> 00:09:44.080
Uh so you just referenced here the CPH 6670.
00:09:44.240 --> 00:09:45.759
Can you talk to me a little bit about that?
00:09:45.840 --> 00:09:47.840
Is it a field or a lab calibrator?
00:09:48.240 --> 00:09:56.320
Uh this is one of our field calibration products and um different than the 6050, this is a lightweight portable device.
00:09:56.399 --> 00:10:01.679
So this makes it a lot easier for our customers to take this directly to the field and calibrate these instruments.
00:10:01.759 --> 00:10:11.120
Um, a few other features that that make this more capable for field calibration is that um for those that are watching, you can see here it's in a um plastic pelican case.
00:10:11.200 --> 00:10:12.720
And so it's it's a lot more rugged.
00:10:12.879 --> 00:10:14.080
Again, lightweight more portable.
00:10:14.159 --> 00:10:17.279
The 60-50 weighs, way too much to be carrying around.
00:10:17.519 --> 00:10:26.240
Um lightweight more portable, and also a
key feature with this is that it also has an internal pump and um um a vacuum pump.
00:10:26.320 --> 00:10:33.600
So it's got pressure and vacuum sources inside that can generate pressures from negative uh 0.9 bar all the way up to 70 bar.
00:10:33.840 --> 00:10:40.159
Um, whereas the if I, you know, 6050, for example, I would actually have to hook up um pressure and vacuum sources to it.
00:10:40.240 --> 00:10:44.879
And so it's gonna need an entire environment around it to operate as um as you need it.
00:10:45.039 --> 00:10:47.039
So this is everything you need is internal.
00:10:47.120 --> 00:10:51.440
Another beautiful thing about this too is that it does have switch testing, which the 6050 does as well.
00:10:51.600 --> 00:11:00.960
And you have um you can do loop calibrations, and this has uh the ability to source and measure um pressure or voltage and um lamps as well.
00:11:01.279 --> 00:11:04.879
So um, you know, it's not meant to be a lab setup.
00:11:05.039 --> 00:11:13.759
This then one of the downsides of this, of course, is you sacrifice um accuracy and certainty with these products uh when you take it to the field.
00:11:14.000 --> 00:11:16.159
Um, but it's it's meant to be practical.
00:11:16.320 --> 00:11:19.759
So again, it's meant to be taken on-site and calibrate these devices where they're at.
00:11:20.080 --> 00:11:20.960
That makes a lot of sense.
00:11:21.039 --> 00:11:25.519
So it's all about kind of weighing like what kind of accuracy versus what kind of convenience works best.
00:11:25.759 --> 00:11:26.639
Yeah, that's pretty cool.
00:11:26.799 --> 00:11:29.600
And do you think that uh this product strikes that balance?
00:11:29.679 --> 00:11:33.440
Like I know it's not as accurate, but do you think it kind of strikes a good balance between the two?
00:11:33.679 --> 00:11:35.120
Yeah, and that's kind of interesting you brought that up.
00:11:35.200 --> 00:11:43.600
So there is sort of a spectrum, so to say, between lab calibrators on one side and and field calibrators on the other.
00:11:43.679 --> 00:11:54.000
You know, on one side with the lab calibrators, you're gonna have um very high uncertainty, um, but low portability, low um efficiency, et cetera.
00:11:54.080 --> 00:12:00.639
And then the field calibrators, you're gonna have those other trade-offs, you know, uh high efficiency, being able to reduce downtime, uh, et cetera.
00:12:00.799 --> 00:12:03.679
But the issue is again, you're sacrificing uncertainty here.
00:12:03.759 --> 00:12:09.600
So, you know, we have products like our CPH 6670 on that far end for portable calibrations.
00:12:09.679 --> 00:12:14.960
We have the CPH 7000, 8000 that um all of them do a few different things, some better than others.
00:12:15.120 --> 00:12:19.679
You have our 6050 on the the right hand side, I would say, for the the lab calibration.
00:12:19.759 --> 00:12:22.480
We also have the CPC 8000, etc.
00:12:22.799 --> 00:12:39.919
Um, two sides of the spectrum, you know, there
is um we have customers all the time asking, hey, um, you know, is there potential to get a field calibrator, you know, left side versus a or with excuse me, a um uncertainty um an accuracy capability of a lab calibrator.
00:12:40.000 --> 00:12:45.200
So we get that lab calibration um um potential in the field.
00:12:45.279 --> 00:12:50.240
And so there's an interesting market there, you know, um there's really I I don't know anything on the market that kind of fits that bill.
00:12:50.399 --> 00:12:56.320
So, you know, maybe there's potential to jump in there one day and see if we can blend these two and and meet in the middle where they're at.
00:12:56.879 --> 00:12:57.360
Great.
00:12:57.600 --> 00:13:04.080
Um, so you kind of mentioned the impact of downtime, like more so the in uh the operational impact of downtime.
00:13:04.159 --> 00:13:13.600
So, how should businesses think about that or take that into consideration when they're choosing between sending equipment to get calibrated in a lab versus just bringing it on site?
00:13:13.840 --> 00:13:14.240
Yeah.
00:13:14.480 --> 00:13:20.080
Um, so I like to think of downtime and it I believe it should be included in the calibration costs.
00:13:20.159 --> 00:13:22.639
I'm not sure if if companies truly look at it that way.
00:13:22.799 --> 00:13:31.200
Um, and the reason why is because again, as I talked about earlier, when when you have to remove a device, um, you you have to take it out of the line.
00:13:31.360 --> 00:13:32.559
That line shut down.
00:13:32.639 --> 00:13:36.000
So, you know, that product's not getting moved out, moved through.
00:13:36.080 --> 00:13:41.279
Um, so you have to consider the the loss of revenue during that downtime for um loss of sales.
00:13:41.440 --> 00:13:44.799
Um, you also need to worry about the you know the logistics again.
00:13:44.960 --> 00:13:54.960
So having to pay to ship it out, you have to pay for the the labor, the text to uninstall it, um, to reinstall it, you have to pay for the the again the calibration itself.
00:13:55.120 --> 00:14:02.159
And so you know the the downtime and how that plays in is you have to consider all these factors um into the true calibration cost.
00:14:02.320 --> 00:14:09.919
So you know, maybe a customer looks at um is like, oh, it it's cheaper to calibrate uh to ship it off.
00:14:10.000 --> 00:14:11.759
It's a couple hundred dollars for a calibration.
00:14:11.840 --> 00:14:20.799
Maybe we ship it off and have it that way um versus paying a couple, you know, um a couple thousand for one of these field calibrators and one offsets the other.
00:14:20.960 --> 00:14:42.480
Um, but no, when you really start adding up the true cost of um the downtime and the cost of calibration, especially if you do this over thousands of sensors in a facility, um being able to take something like this directly to them and um getting rid of the downtime and in the other um effects that that um drive costs, um, this would be a much better option in those cases.
00:14:42.720 --> 00:14:42.960
Okay.
00:14:43.120 --> 00:14:43.840
That makes a lot of sense.
00:14:43.919 --> 00:14:49.200
So it's kind of it really is like a case-by-case depending on your needs, your environment, all the different factors.
00:14:49.440 --> 00:14:49.919
Absolutely.
00:14:50.159 --> 00:14:50.879
Um awesome.
00:14:51.120 --> 00:14:58.879
Well, as we wrap up, I was hoping you could kind of just sum up some of the main takeaways between the differences and uh making the choice.
00:14:58.960 --> 00:15:05.039
Um, and if there's anything else you want to say about the new CPH 6670, I'll let you take it away.
00:15:05.279 --> 00:15:05.519
Yeah.
00:15:05.600 --> 00:15:22.639
Um, you know, just kind of in summary, um I like to think about, um, especially when I'm looking at these two and maybe even trying to pick um which one's gonna be right for my um from uh what calibrating I need to do in my my factory, my my facility lab, et cetera.
00:15:22.879 --> 00:15:26.320
Um I like to think about kind of three different things really.
00:15:26.399 --> 00:15:32.320
The one of the first things to look up as an application specialist, I want to know what their accuracy is of the devices they're calibrating.
00:15:32.480 --> 00:15:39.679
So um typically something we we do here in the industry, the the standard what we is uh we call like the four to one rule.
00:15:39.759 --> 00:15:50.080
So what that is is we try to have a primary standard or reference standard that is um has an uncertainty that is four times better than the device that we are um calibrating.
00:15:50.159 --> 00:16:07.600
And so if the the customer has a very um high um uncertainty device, or I guess low uncertainty in that case, high accuracy device um that this might not meet the needs of, you know, of course we're gonna have to push them to um you know something like a lab calibrator that's gonna help meet those standards.
00:16:07.679 --> 00:16:13.600
Um so determine what the actions, if they're lower accuracies, um, you know, then your options are kind of open.
00:16:13.679 --> 00:16:15.039
Maybe you could do one or the other.
00:16:15.360 --> 00:16:30.960
Another thing I like to look at too is again the the downtime and logistic piece, you know, if I take this instrument out of the process, it does that severely impact the um the process line and in the manufacturing of goods or um whatever you're doing with that instrument.
00:16:31.120 --> 00:16:37.519
So if that answer is yes, then um field calibrator CPH 6670 is going to be uh the ticket for you.
00:16:37.679 --> 00:16:44.879
And then the other um little piece I like to look at um would be um you know your scope of work.
00:16:45.120 --> 00:16:49.600
And so um exactly um what are you doing with these sensors?
00:16:49.759 --> 00:16:54.559
Um so you know, a good example, maybe like an application for a CPH 6670.
00:16:54.639 --> 00:17:01.519
Um, you're we're in an oil and gas field, um, you
know, they have a pipeline, um, removing that sensor, a few things to consider.
00:17:01.600 --> 00:17:07.359
One, maybe it's a you know again low uncertainty sensor, it's moving fluid through it so it's contaminated.
00:17:07.519 --> 00:17:13.839
Um, and removing that sensor is also going to shut down that pipeline, huge loss in revenue, et cetera.
00:17:13.920 --> 00:17:19.920
So it's it makes a ton more sense based on that application scope of work to use this CPH 6670.
00:17:20.160 --> 00:17:23.039
Whereas let's do let's do aerospace, for example.
00:17:23.119 --> 00:17:30.480
Um, their six-pack, which is their their air data gauges that pilots will use to determine you know, altitude, airspeed, um, etc.
00:17:30.799 --> 00:17:37.039
Um, in that case, those are usually very, very clean instruments, they're just seeing thematic um air pressure.
00:17:37.359 --> 00:17:40.400
And um, the uncertainty of those instruments are crucial.
00:17:40.480 --> 00:17:48.640
You know, we need to make sure those pilots are um getting very accurate readings so you and I stay safe as we as we fly through the clouds and then those spaceships, right?
00:17:49.119 --> 00:17:58.400
And so um in that case, you know, they're gonna need a lab calibrator with that that level of um uncertainty to calibrate those devices.
00:17:58.559 --> 00:17:58.880
Okay.
00:17:59.200 --> 00:18:02.160
And then, yeah, so that's how I kind of look at picking between the two.
00:18:02.240 --> 00:18:10.079
And again, in summary, um, if I'm looking for something with utmost confidence and and highest level of uncertainty, I'm looking at a um lab calibrator.
00:18:10.160 --> 00:18:14.400
And if I need something more practical, if I'm worried about efficiency, I'm gonna look for something like a field calibrator.
00:18:14.559 --> 00:18:17.680
You know, I just want to say too, um, first off, thank you for having me on.
00:18:17.759 --> 00:18:20.720
I know this is uh season three, episode one, I guess I'm the lucky first guest.
00:18:20.960 --> 00:18:21.920
You're the lucky first guest.
00:18:22.160 --> 00:18:22.880
Super exciting.
00:18:23.039 --> 00:18:28.000
And um, it's been a blast kind of talking about these two products and and things that we have going on here at Menzer.
00:18:28.160 --> 00:18:34.559
And um, again, I'm the application specialist here, so I'd love to hear what um all of our customers are doing out there in the field.
00:18:34.640 --> 00:18:45.599
And should they have any other questions or um want to inquire more about our products or just want to talk about applications and what um and get them steered in the right direction or to bounce ideas off myself or some of our engineers here.
00:18:45.759 --> 00:18:47.440
Um I'd love for them to reach out to us.
00:18:47.519 --> 00:18:53.920
Um, they can reach us at sales at menser.com or they can call us here if they want to talk to a real person, which is pretty cool.
00:18:54.000 --> 00:18:55.359
Um, not a not a robot.
00:18:55.440 --> 00:18:57.200
Um they can call us here in San Marcus.
00:18:57.279 --> 00:19:06.000
You know, our number is 512-396-4200, and we can get you a hold of somebody, um maybe myself, and um love to talk pressure with you.
00:19:06.240 --> 00:19:06.559
Awesome.
00:19:06.720 --> 00:19:07.519
Well, thanks so much.
00:19:07.599 --> 00:19:11.839
We really enjoyed having you on and excited to see uh the rest of the season.
00:19:12.000 --> 00:19:15.599
We were lucky enough to have you as our first guest for season three.
00:19:15.839 --> 00:19:19.599
Um, so thank you so much for watching, and we'll be back with our second episode.
00:19:19.839 --> 00:19:20.079
Let's go.
00:19:20.240 --> 00:19:21.039
Eat them up cats, right?
00:19:21.200 --> 00:19:22.160
Eat them up cats.
00:19:23.039 --> 00:19:24.000
Thank you for joining us.
00:19:24.160 --> 00:19:25.839
We hope you enjoyed today's podcast.
00:19:26.000 --> 00:19:30.799
To listen to more topics, you can visit info.menster.com forward slash podcast.
00:19:30.960 --> 00:19:36.480
And if there's a topic you're interested in hearing, leave a comment on our YouTube channel or reach out on any of our social platforms.
00:19:36.640 --> 00:19:37.359
Thanks for listening.
00:19:37.519 --> 00:19:39.440
You stay classy metrology lovers.
00:19:39.920 --> 00:19:40.559
Crushed it.
00:19:40.720 --> 00:19:41.920
I feel like we crushed it, right?
00:19:42.559 --> 00:19:43.599
Well, exactly.